Vista de pagina 19
20
Lamp Replacement Instructions
WARNING: ONLY A QUALIFIED SERVICE TECHNICIAN MAY REPLACE THE SYSTEM LAMP.
PROTECTIVE EYEWEAR MUST BE WORN. NEVER POINT THE WINDOW OF THE LAMP TO
UNPROTECTED TISSUE.
Symptoms indicating need for replacement are:
1. Low intensity on system light meter.
2. Lamp fails while in use or does not turn on.
To Remove the Old Lamp
1. Turn unit off and unplug the power cord from its power source.
2. Wait for the unit and the lamp to cool.
3. Unscrew thumbscrews to remove the lamp access plate.
4. Grasp lamp and slide towards you to remove from light source.
5. Unplug the lamp from the 2-pin lamp socket.
To Insert New Lamp
1. Orient the lamp facing forward in the lamp guides on the chassis.
2. Push toward the top of the unit until the lamp slides securely onto the lamp pins.
3. Replace the lamp access plate and tighten thumbscrews.
NOTE: Make sure the access plate fits correctly in the cutout provided. All corners should be even with the
surrounding surfaces. If the access plate is misaligned, the unit may not receive AC power.
Unscrew thumbscrews
to remove access door.
HANDLES BRASS SOCKETS
Vista de pagina 19